Does the use of the largest possible metal head increase the wear of vitamin E-doped cross-linked polyethylene? Two-year results from a randomized controlled trial.

Department of Orthopaedics, Sahlgrenska University Hospital, Gothenburg, Sweden. Department of Orthopaedics, University of Gothenburg, Sahlgrenska Academy, Institute of Clinical Sciences, Gothenburg, Sweden. Department of Clinical Medicine, Faculty of Health Sciences, University of Copenhagen, Copenhagen, Denmark. Department of Orthopaedic Surgery, Copenhagen University Hospital, Hvidovre, Denmark. Harris Orthopaedic Laboratory, Massachusetts General Hospital, Boston, Massachusetts, USA. Harvard Medical School, Harvard University, Boston, Massachusetts, USA.

The bone & joint journal. 2021;(7):1206-1214

Abstract

AIMS: We aimed to investigate if the use of the largest possible cobalt-chromium head articulating with polyethylene acetabular inserts would increase the in vivo wear rate in total hip arthroplasty. METHODS In a single-blinded randomized controlled trial, 96 patients (43 females), at a median age of 63 years (interquartile range (IQR) 57 to 69), were allocated to receive either the largest possible modular femoral head (36 mm to 44 mm) in the thinnest possible insert or a standard 32 mm head. All patients received a vitamin E-doped cross-linked polyethylene insert and a cobalt-chromium head. The primary outcome was proximal head penetration measured with radiostereometric analysis (RSA) at two years. Secondary outcomes were volumetric wear, periacetabular radiolucencies, and patient-reported outcomes. RESULTS At two years, 44 patients in each group were available for RSA assessment. The median total two-year proximal head penetration was -0.02 mm (IQR -0.09 to 0.07; p = 0.548) for the largest possible head and -0.01 mm (IQR -0.07 to 0.10; p = 0.525) for 32 mm heads. Their difference was not statistically significant (p = 0.323). Neither group demonstrated a bedding-in period. The median steady-state volumetric wear rates were 6.1 mm3/year (IQR -59 to 57) and 3.5 mm3/year (-21 to 34) respectively, and did not differ between the groups (p = 0.848). There were no statistically significant differences in periacetabular radiolucencies or patient-reported outcomes. CONCLUSION The use of the largest possible metal head did not increase vitamin E-doped cross-linked polyethylene wear compared with 32 mm heads at two years. Linear wear was negligible and volumetric wear rates were very low in both head size groups. There was a tendency towards higher values of volumetric wear in large heads that warrants longer-term evaluation before any definite conclusions about the association between head size and wear can be drawn. Cite this article: Bone Joint J 2021;103-B(7):1206-1214.

Methodological quality

Metadata